Cleaning Guide for Schiller MS-12 ECG Machine
- Do not immerse the device or accessories in liquid.
- Do not sterilise using hot water, steam, hot air, or an
autoclave.
- Avoid cleaning products that contain strong acids or
chemicals that may damage plastic.
- Always follow the manufacturer’s dilution instructions for
cleaning solutions.
- Never use ethyl alcohol, ethanol, acetone, hexane, or
abrasive cleaners.
- Do not spray cleaning solution directly onto the device.
- Ensure no liquid enters connectors or openings.
- Avoid pulling or stressing patient cables.
- Wearing protective gloves is recommended.
General Cleaning Instructions
Clean the MS-12 USB ECG Recorder and its cables with a damp,
not wet, cloth. A mild cleaner or 50% isopropyl alcohol may be used for
fingerprints or grease.
MS-12 Unit
- Wipe the device with a slightly damp cloth using a cleaning solution approved for use.
- Remove any extra liquid to prevent pooling.
- Make sure liquid does not enter any connectors or seams.
- If liquid gets inside a connector, dry it with warm air
and check that the device works properly afterwards.
Cleaning Cable Assemblies
- Check the cable for damage before cleaning.
- Wipe the cable using a cloth slightly moistened with an
approved cleaner.
- Clean the cable in short 20 cm sections to avoid bending
or bunching the insulation.
- Avoid letting liquid enter the connectors.
- If liquid enters a connector, dry gently with warm air.
Approved Cleaning Solutions
- 50% isopropyl alcohol
- Mild neutral detergent
- Plastic-safe cleaning products
Solutions That Must NOT Be Used
- Ethyl alcohol
- Acetone
- Hexane
- Abrasive powders or cleaners
- Products that can dissolve plastic
Disinfection Instructions
Use disinfectants suitable for clinical environments. Apply
disinfectant using the same method as cleaning—using a damp cloth and ensuring
no liquid enters the device.
Approved Disinfectants
- Isopropyl alcohol (50%)
- Propanol (35%)
- Ethyl hexanal
- Aldehyde (2–4%)
- Ethanol (50%)
- Products are safe for PC/ABS plastic
Recommended products by the manufacturer
- Bacillol 30 Tissues
- Bacillol 30 Foam
- Mikrozid liquid
- Mikrozid wipes
Disinfectants That Must NOT Be Used
- Organic solvents
- Ammonia-based detergents
- Abrasive cleaning agents
- 100% alcohol, Virex, Sani‑Master
- Sani-Cloth, Ascepti, Clorox wipes
- HB Quat
- Household cleaners (Fantastic, Tilex)
- Conductive solutions
- Acetone, ammonium chloride, Betadine, chlorine/wax
products, ketone, sodium salts