How to perform data reset
How to Perform a Data Reset on the NObreath® Monitor
1. Access the Data Reset Button
- Remove the back cover and the breath drying cartridge to expose the Data Reset button at the top right of the NObreath®.
- Use the provided screwdriver to unscrew the screw on the back of the monitor.
- Slide and lift off the back cover to fully access the reset button.
2. Perform the Data Reset
- Press and hold the Data Reset button for 5 seconds until the reset icon appears on the screen.
- Important: A data reset will erase all patient data and reset the PIN to the default (0000).
3. Confirm or Cancel the Reset
- Press the tick (✔) button to confirm.
- Press the cross (✖) button to cancel.
- An hourglass icon will appear while the device erases all data.
- ⏳ This process may take up to 5 minutes to complete.