How to perform data reset

How to perform data reset

How to Perform a Data Reset on the NObreath® Monitor

1. Access the Data Reset Button

  • Remove the back cover and the breath drying cartridge to expose the Data Reset button at the top right of the NObreath®.
  • Use the provided screwdriver to unscrew the screw on the back of the monitor.
  • Slide and lift off the back cover to fully access the reset button.

2. Perform the Data Reset

  • Press and hold the Data Reset button for 5 seconds until the reset icon appears on the screen.
  • Important: A data reset will erase all patient data and reset the PIN to the default (0000).

3. Confirm or Cancel the Reset

  • Press the tick (✔) button to confirm.
  • Press the cross (✖) button to cancel.
  • An hourglass icon will appear while the device erases all data.
  • This process may take up to 5 minutes to complete.