Results constantly 5 or 6 ppb

Results constantly 5 or 6 ppb

Please check the data to confirm whether all values are consistently showing 5 and 6. 
  1. If confirmed, this may indicate a fault with the NO sensor.
Next steps:
  1. If no alternative NO sensor is available:
    1. Reinsert the current NO sensor.
    2. Run the test again.
    3. If the test values return to normal, continue using the sensor.
    4. If the issue persists, the sensor should be replaced.
  2. If a new or verified working NO sensor is available:
    1. Replace the current sensor with the new one.