Write XML Error: Could not find part of the path
If you encounter a 'Write XML Error: Could not find part of the path' error when attempting to use your Universal Smart ECG machine, follow the instructions below to resolve this issue.

or

This guide can also be used by
customers with a Universal USB ECG machine.
- Open CardioView from the desktop

- Click on ECG SETTINGS

- Set the Measurement System to Metric

- On the Analysis tab, some surgeries prefer to change the QTc
Calculation to Fridericia. This is a clinical decision, and we would recommend
speaking with your GP or consultant if you are unsure. This will affect how the
software measures QTc on the reports. Bazett is the default but older
calculation; some clinicians prefer Fridericia, which is a newer method.

- Check the highlighted parts. The power
filter must be set to 50 Hz for UK and European use, then click OK.

- Click PRINT SETTINGS

- Typically, you will only need to change the
highlighted part and click OK once done
] - Close down CardioView
- Run the ECG again via the Intelligent Integration Interface.
If you continue to experience issues please contact Numed support via the Chat button on this page.