Sunvou FeNO
Servicing and Calibrating your Sunvou FeNO Monitor
The Sunvou FeNO monitor does not require annual servicing as the sensor is replaced once your 100, 300 or 500 tests pack has been completed. The device performs an automatic purge cycle after each patient test; during this, it will draw ambient air ...
Results constantly 5 or 6 ppb
Please check the data to confirm whether all values are consistently showing 5 and 6. If confirmed, this may indicate a fault with the NO sensor. Next steps: If no alternative NO sensor is available: Reinsert the current NO sensor. Run the test ...
Cleaning and Disinfecting the Sunvou FeNO Monitor
Each test is performed using a disposable bacterial and viral filtered mouthpiece, which is 99.99% efficient; there is minimal risk of cross-infection. A replacement handpiece is supplied with every 100 mouthpieces. Routine Cleaning To clean all ...
Sunvou FeNO Monitor Error Codes
Code Cause Solution E01 Device self-checking failed, please restart Restart the device E02 NO sensor not connected Reinstall the sensor correctly. E03 NO tests depleted The number of remaining measurements is displayed at the top of the screen. If ...
Changing the Shutdown Time
To change the length of time it takes before your Sunvou FeNO monitor will automatically shutdown, go to the settings cog and click on Authorisation Code The code to change the shutdown time is 835736
Changing the Time and Date
To change the time and date on your Sunvou FeNO monitor, go to the settings cog and click on Authorisation Code The code to change the time and date is 080228